PIE Medical Imaging B.V. recalls mensio Workstation (Vascular Fenestrated) software

Published
29 Jul 2026
Units affected
165
Agency
FDA
Country
United States

The hazard

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

3mensio Workstation (Vascular Fenestrated) software

Brand / company
PIE Medical Imaging B.V.
Codes / lots
Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
Distribution
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
Recall status
Ongoing