PIE Medical Imaging B.V. recalls mensio Workstation (Vascular…
PIE Medical Imaging B.V. recalls mensio Workstation (Vascular Fenestrated) software
- Published
- 29 Jul 2026
- Units affected
- 165
- Agency
- FDA
- Country
- United States
The hazard
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
3mensio Workstation (Vascular Fenestrated) software
- Brand / company
- PIE Medical Imaging B.V.
- Codes / lots
- Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 29 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.