CheckMyRecalls

PFM MEDICAL INC. recalls Multi-Snare Set

Published
1 Jul 2026
Units affected
32
Agency
FDA
Country
United States

The hazard

A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

Brand / company
PFM MEDICAL INC.
Codes / lots
REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)
Distribution
US Nationwide distribution in the states of NY, MD.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.