Permobil Inc recalls M5 Corpus Power Wheelchair REF,120032
- Published
- 5 Aug 2026
- Units affected
- 715 wheelchairs
- Agency
- FDA
- Country
- United States
The hazard
This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
M5 Corpus Power Wheelchair REF,120032
- Brand / company
- Permobil Inc
- Codes / lots
- Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
- Distribution
- Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 5 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.