CheckMyRecalls

Permobil Inc recalls M5 Corpus Power Wheelchair REF,120032

Published
5 Aug 2026
Units affected
715 wheelchairs
Agency
FDA
Country
United States

The hazard

This recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

M5 Corpus Power Wheelchair REF,120032

Brand / company
Permobil Inc
Codes / lots
Power Platform firmware versions 1.27 through 1.57 UDI-DI 17330818334562
Distribution
Worldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 5 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.