CheckMyRecalls

AVID Medical, Inc. recalls Halyard convenience kits containing…

AVID Medical, Inc. recalls Halyard convenience kits containing recalled Medline Namic Angiogra…

Published
12 Aug 2026
Units affected
3 products
Agency
FDA
Country
United States

The hazard

Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05

Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05

Brand / company
AVID Medical, Inc.
Codes / lots
UDI-DI 10809160388495 Lots 1561685 1575926 1583048 1584017 1589652 1595715 1599634 1605749 1611810 1621476 1637222 1637224 1637225 1643118 1645143 1645637 1663929 1663930 1663931 1663932 | UDI-DI 10809160377314 Lots 1573593 1583827 1609541 1612134 1612516 1619132 1623384 1627013 1635459 1672306 1675159 | UDI-DI 10809160408650 Lots 1618351 1621441 1627930 1635908 1644501 1646021 1651957 1654887 1656581 1656582 1659713 1670884 1670885 1671950
Distribution
US Nationwide distribution in the states of NC and TX.
Recall status
Ongoing