Erbe USA Inc recalls ERBECRYO 2 Cryosurgical Unit Model/Catalog Number

Published
19 Aug 2026
Units affected
1126 units
Agency
FDA
Country
United States

The hazard

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Brand / company
Erbe USA Inc
Codes / lots
UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
Distribution
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
Recall status
Ongoing