CheckMyRecalls

Fort Defiance Industries, LLC recalls Fort Defiance Industries (FDI)

Published
26 Aug 2026
Agency
FDA
Country
United States

The hazard

The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Brand / company
Fort Defiance Industries, LLC
Codes / lots
Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 26 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.