CheckMyRecalls

Nova Biomedical Corporation recalls StatStrip Glucose and B- Ketone…

Nova Biomedical Corporation recalls StatStrip Glucose and B- Ketone meter and 1 more

Published
23 Sept 2026
Units affected
5 products
Agency
FDA
Country
United States

The hazard

A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is scanned, resulting in the prior barcode information remaining displayed on the screen for the current test. When this defect occurs and the operator accepts the incorrectly displayed patient information without recognizing the mismatch, the current patient's glucose result will be incorrectly assigned to the previous patient's medical record and transmitted to the data management system (DMS), and the current patient's medical record will not receive a result, leaving their glucose status unrecorded and their clinical management without the information needed to guide treatment decisions. The previous patient, whose medical record now contains a glucose result that does not belong to them, may receive inappropriate clinical intervention based on that erroneous result, particularly in critically ill patients, neonates, and infants who cannot communicate symptoms or self-treat. The defect also affects QC barcode scanning: if a QC scan triggers the same retention behavior, QC results may be attributed to an incorrect identifier or QC tests may fail without the user recognizing the underlying cause. This defect does not affect the analytical accuracy of glucose measurements.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

StatStrip Glucose and B- Ketone meter. Model No: 63910. Glucose test system.

StatStrip Glucose and B- Ketone meter. Model No: 63683. Glucose test system.

StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.

StatStrip Glucose and B- Ketone meter. Model No: 66429. Glucose test system.

StatStrip Glucose Hospital Meter System. Model No: 66739. Glucose test system.

Brand / company
Nova Biomedical Corporation
Codes / lots
Model No: 63910. UDI-DI: 00385480639101. Software Versions: All versions 0.00.13.10 through 1.5.14.25. | Model No: 63683. UDI-DI: 00385480636834. Software Versions: All versions 0.00.13.10 through 1.5.14.25. | Model No: 63685. UDI-DI: 00385480636858. Software Versions: All versions 0.00.13.10 through 1.5.14.25. | Model No: 66429. UDI-DI: :00385480664295. Software Versions: All versions 0.00.13.10 through 1.5.14.25. | Model No: 66739. UDI-DI: 0038548063685. Software Versions: All versions 0.00.13.10 through 1.5.14.25.
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Austria, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, Croatia, Dominican Rep., Finland, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Korea, Switzerland, Thailand, Turkey, United Kingdom.