Fresenius Medical Care Renal Therapies Group, LLC recalls X HD/HDF…

Fresenius Medical Care Renal Therapies Group, LLC recalls X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8 and 3 more

Published
9 Sept 2026
Units affected
4 products
Agency
FDA
Country
United States

The hazard

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

5008X HD/HDF with Twister Blood Tubing Set Part Number 03-5350-8

5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6

5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3

5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C

Brand / company
Fresenius Medical Care Renal Therapies Group, LLC
Codes / lots
Lots 25JR01802 25NR01017 25NR01266 UDI-DI 00840861102525 | Lots 25KR01227 25NR01257 UDI-DI 00840861102556 | Lots 25JR01800 25KR01224 25LR01001 25NR01018 25NR01272 25PR01077 25PR01219 25PR01220 UDI-DI 00840861102518 | Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Recall status
Ongoing