Beckman Coulter Inc. recalls DxC 500i Clinical Analyzer Modules and 1…
Beckman Coulter Inc. recalls DxC 500i Clinical Analyzer Modules and 1 more
- Published
- 14 Oct 2026
- Units affected
- 2 products
- Agency
- FDA
- Country
- United States
The hazard
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
- Brand / company
- Beckman Coulter Inc.
- Codes / lots
- Software version 1.3.5 or earlier. 1. UDI-DI 14987666545089 2. UDI-DI 14987666545072 3. UDI-DI 15099590742331 | Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
- Distribution
- Worldwide - US Nationwide distribution.
- Recall status
- Ongoing