SI-BONE, Inc. recalls iFuse TORQ Implant System Model/Catalog Number

Published
21 Oct 2026
Units affected
100 units
Agency
FDA
Country
United States

The hazard

The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)

Brand / company
SI-BONE, Inc.
Codes / lots
10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911
Distribution
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Recall status
Ongoing