Boston Scientific Cork, Ltd. recalls Direxion Torqueable Microcatheter
- Published
- 28 Oct 2026
- Units affected
- 41 products
- Agency
- FDA
- Country
- United States
The hazard
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/155; Material Number: M001195750
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/130; Material Number: M001195840
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/130; Material Number: M001195240
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/130; Material Number: M001195340
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/STRAIGHT/1RO/130; Material Number: M001195810
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/105; Material Number: M001195230
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/155; Material Number: M001195450
- Brand / company
- Boston Scientific Cork, Ltd.
- Codes / lots
- Material Number: M001195290; UDI-DI: 08714729839408; Batch Numbers: 38362172, 38362173, 38530450, 38595659; | Material Number: M001195440; UDI-DI: 08714729839514; Batch Numbers: 38181961, 38198832, 38224582, 38259483, 38324070, 38460469, 38560455, 38560456, 38595654, 38649857, 38745194, 38745195, 38810586, 38810591, 38904805, 39092267, 39092272, 39143222, 40470259, 40478976, 40478977, 40492318, 40497671, 40497858, 40502038, 40502039; | Material Number: M001195750; UDI-DI: 08714729839729; Batch Numbers: 38049562, 38112644, 38112645, 38181975, 38259489, 38273638, 38342770, 38346223, 38399380, 38707974, 39074174, 39074176, 39092273, 40470667, 40476303, 40485786, 40485787, 40493318, 40495100, 40507439, 40512986, 40512987, 40523966, 40524003, 40531398, 40531399; | Material Number: M001195280; UDI-DI: 08714729839392; Batch Numbers: 38149679, 38323401, 38444072, 38506482, 38506491, 38560458, 38595645, 38810252, 38936816, 39140997, 39141264, 39189785, 39189791, 40482687, 40493382, 40512909; | Material Number: M001195260; UDI-DI: 08714729839378; Batch Numbers: 37818211, 37881104, 37942459, 38002443, 38224536, 38415913, 38595894, 38595895, 38629331, 38687224, 38904567, 38904568, 38916620, 38
- Distribution
- Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 28 Oct 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.