Baxter Healthcare Corporation recalls TruSystem 7500 Hybrid Plus (FC)…
Baxter Healthcare Corporation recalls TruSystem 7500 Hybrid Plus (FC) and 4 more
- Published
- 11 Nov 2026
- Units affected
- 5 products
- Agency
- FDA
- Country
- United States
The hazard
There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087
Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
TruSystem 7500 surgical table, Model/Catalog Number: 4091000
- Brand / company
- Baxter Healthcare Corporation
- Codes / lots
- Serial numbers: 107971925; UDI-DI number: 00887761974067 | Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519 | Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776,
- Distribution
- Worldwide - US Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Nov 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.