Baxter Healthcare Corporation recalls TruSystem 7000 U Operating Table…
Baxter Healthcare Corporation recalls TruSystem 7000 U Operating Table (dV) and 1 more
- Published
- 11 Nov 2026
- Units affected
- 2 products
- Agency
- FDA
- Country
- United States
The hazard
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
TruSystem 7000 U Operating Table (dV), Product Code 1723633
Mobile column TruSystem 7500 U, Product Code 1730720
- Brand / company
- Baxter Healthcare Corporation
- Codes / lots
- UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662 | UDI-DI: 00887761974388; Serial number: 110183858
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 11 Nov 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.