CheckMyRecalls

Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The…

Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer…

Published
1 Jan 2025
Units affected
7295 programmers
Agency
FDA
Country
United States

The hazard

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Brand / company
Abbott Medical
Codes / lots
Programmer Model Number: 3650 UDI-DI code: 05414734509725 Software Version Number / Serial Numbers: 28.0.1 REV 2 12995 28.0.1 REV 2 16103 28.0.1 REV 2 30831 28.0.1 REV 2 12040778 28.0.1 REV 2 12041825 28.0.1 REV 2 12048943 28.0.1 REV 2 12050154 28.0.1 REV 2 12051352 28.0.1 REV 2 12051615 28.0.1 REV 2 12056133 28.0.1 REV 2 12713 28.0.1 REV 2 14655 28.0.1 REV 2 23324 28.0.1 REV 2 24050 28.0.1 REV 2 12042174 28.0.1 REV 2 12042239 28.0.1 REV 2 12048730 28.0.1 REV 2 11959 28.0.1 REV 2 16116 28.0.1 REV 2 16571 28.0.1 REV 2 27499 28.0.1 REV 2 12056125 28.0.1 REV 2 15378 28.0.1 REV 2 18041 28.0.1 REV 2 27164 28.0.1 REV 2 12045425 28.0.1 REV 2 12045483 28.0.1 REV 2 12870 28.0.1 REV 2 14910 28.0.1 REV 2 18143 28.0.1 REV 2 21904 28.0.1 REV 2 23317 28.0.1 REV 2 12043799 28.0.1 REV 2 12044806 28.0.1 REV 2 12055973 28.0.1 REV 2 24842 28.0.1 REV 2 12040739 28.0.1 REV 2 12042686 28.0.1 REV 2 12044212 28.0.1 REV 2 12045527 28.0.1 REV 2 12052141 28.0.1 REV 2 12055384 28.0.1 REV 2 12052537 28.0.1 REV 2 19660 28.0.1 REV 2 21099 28.0.1 REV 2 24610 28.0.1 REV 2 30022 28.0.1 REV 2 12044540 28.0.1 REV 2 12047076 28.0.1 REV 2 12053304 28.0.1 REV 2 12706 28.0.1 REV 2 20207 28.0.1 REV 2 20933 28.0.1 REV 2 22
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN TX, UT, VA, WA, WI, and WV. The countries of Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 1 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.