CheckMyRecalls

Abbott Medical recalls Aveir DR Leadless Pacemaker (LP) Ventricular…

Abbott Medical recalls Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI…

Published
23 Apr 2025
Units affected
6 devices
Agency
FDA
Country
United States

The hazard

Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Aveir DR Leadless Pacemaker (LP) Ventricular Device REF LSP202V UDI-DI code: 05415067040725 REF LSP112V UDI-DI code: 05415067034472 The Aveir leadless pacemaker system supports the implantation and use of an Aveir Leadless Pacemaker (LP) within the targeted chamber(s) of the heart for monitoring a patient s heart rate and providing rate responsive bradycardia pacing therapy to regulate heart rate.

Brand / company
Abbott Medical
Codes / lots
Model Number: LSP202V UDI-DI code: 05415067040725 Serial Numbers: 1357831, 1357869, 1357885, 1357894, 1357895 Model Number: LSP112V UDI-DI code: 05415067034472 Serial Number: 1357956
Distribution
Worldwide - U.S. Nationwide distribution in the states of FL, SC, and WV and the country of Spain. There were five (5) LSP202V devices distributed within the United States (3 in FL, 1 in SC, 1 in WV) that are impacted by this issue, two (2) of which were implanted, see Attachment A.3 for list of serial numbers. There is 1 LSP112V device that is impacted by this issue, but this was distributed outside of the United States. See Attachment A.4 for list of OUS serial numbers.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 23 Apr 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.