CheckMyRecalls

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare…

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Brivo NM 615 and 12 more

Published
22 Jan 2025
Units affected
13 products
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

GE Healthcare Brivo NM 615, Model/Catalog Numbers: 1) 5376204-15, 2) H2401MA, 3) H3100TA; System, Tomography, Computed, Emission

GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography, Computed, Emission

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

GE Healthcare Discovery NM/CT 670 DR, Model/Catalog Numbers: 1) 5376204-70-56-2; 2) H2401AW; 3) H3100AN; 4) H3100BT; 5) H3100PX; System, Tomography, Computed, Emission

GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission

GE Healthcare Discovery NM/CT 670, Model/Catalog Numbers: 1) 5376204-70-54-2, 2) H2401BC, 3) H2401CE, 4) H2401CG, 5) H3100AG, 6) H3100JW, 7) H3100NC; System, Tomography, Computed, Emission

GE Healthcare Infinia I Hawkeye 1, Model/Catalog Numbers: 1) H3000WN; 2) H3000YM; 3) H3000YS; System, Tomography, Computed, Emission

GE Healthcare Infinia II, Model/Catalog Numbers: 1) H2799ZP; 2) H3000WT; System, Tomography, Computed, Emission

GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission

GE Healthcare Infinia II Hawkeye 4, Model/Catalog Numbers: 1) H2799PP; 2) H2799PR; 3) H2799YC; 4) H2799YD; 5) H3000WC; 6) H3000WD; 7) H3000WT; 8) H3000WW; 9) H3000WY; 10) H3000WZ; System, Tomography, Computed, Emission

GE Healthcare Discovery NM 630, Model/Catalog Numbers: 1) 5376204-30-2, 2) H2401CF, 3) H2401CH, 4) H2401JE, 5) H3100JY; System, Tomography, Computed, Emission

GE Healthcare Discovery NM/CT 670 PRO, Model/Catalog Numbers: 1) 5376204-70-54-2; 2) H2401AK; 3) H2402AC; 4) H3100AH; 5) H3100XB; System, Tomography, Computed, Emission

Brand / company
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Codes / lots
1) 5376204-15: UDI/DI 00840682145503, System ID 502629NMGD, Serial Number NT2UG2100002BH; UDI/DI 00840682117869, System ID 509455BRIVO, Serial Number NT24G1700010BH, UDI/DI N/A, System ID 248544D630, Serial Number NT2QG1900003BH\NT2UG2100001BH. 2) H2401MA: UDI/DI Not Available, System ID 501552D615, Serial Number NT24G1600023; UDI/DI Not Available, System ID 100035454149, Serial Number T2CG15015/100035454149; UDI/DI Not Available, System ID 337643NM, Serial Number NT24G1600021; UDI/DI Not Available, System ID 718963B615, Serial Number NT24G1700009; UDI/DI Not Available, System ID Not Available, Serial Number NT24G1600020; UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1700010; UDI/DI Not Available, System ID 509455BRIVO, Serial Number NT24G1800001; UDI/DI (01)00840682117869(11)17060021NT24G1700003BH, System ID NM6583AR, Serial Number NT24G1700003; UDI/DI (01)00840682117869(11)17090021NT24G1700006BH, System ID NT24G1700006BH, Serial Number NT24G1700006; UDI/DI (01)00840682117869(11)17080021NT24G1700004BH, System ID NT24G1700004BH, Serial Number NT24G1700004; UDI/DI Not Available, System ID Not Available, Serial Number NT24G1700008; UDI/DI Not Available, System ID CS
Distribution
Worldwide
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 22 Jan 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.