The hazard

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

SMV DSI, System, Tomography, Computed, Emission

SMV BodyTrack, System, Tomography, Computed, Emission

SMV DSX New Line, System, Tomography, Computed, Emission

MAXXUS, System, Tomography, Computed, Emission

Brand / company
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Codes / lots
UDI/DI Not applicable: All serial numbers in distribution | UDI/DI Not applicable: All serial numbers in distribution | UDI/DI Not applicable: All serial numbers in distribution | UDI/DI Not applicable: All serial numbers in distribution
Distribution
US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 21 Jan 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.