GE Medical Systems, LLC recalls Optima XR646 HD and 3 more
- Published
- 12 Feb 2025
- Units affected
- 4 products
- Agency
- FDA
- Country
- United States
The hazard
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Optima XR646 HD. X-Ray imaging system.
Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
Discovery XR656 HD. X-Ray imaging system.
Discovery XR656HD. X-Ray imaging system.
- Brand / company
- GE Medical Systems, LLC
- Codes / lots
- GTIN: 00195278526892 | GTINs: 00840682138598, 00195278840226, 00840682136709) | Definium Tempo (GTIN: 00195278118356) Definium Tempo Pro (GTIN: 00195278070265) | GTINs: 00840682142175, 00840682142168
- Distribution
- Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.