GE Medical Systems, LLC recalls GE Healthcare Revolution Apex and 6 more
- Published
- 19 Mar 2025
- Units affected
- 7 products
- Agency
- FDA
- Country
- United States
The hazard
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed
GE Healthcare Revolution CT, System, X-ray, Tomography, Computed
GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
GE Healthcare Revolution Apex Expert, System, X-ray, Tomography, Computed
GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
- Brand / company
- GE Medical Systems, LLC
- Codes / lots
- System ID Number: 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number: 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number: 630266CTRV, UDI (01)00195278460271(11)240800(21)REV2X2400093CN, Mfg. Lot/Serial Number REV2X2400093CN; System ID Number: 773702REVCT6ED, UDI (01)00195278460271(11)231000(21)REV2X2300071CN, Mfg. Lot/Serial Number REV2X2300071CN; System ID Number: 312996CTRV2, UDI (01)00195278460271(11)241000(21)REV2X2400126CN, Mfg. Lot/Serial Number REV2X2400126CN; System ID Number: 1000000018, UDI To be provided, Mfg. Lot/Serial Number REV2X2300074CN; System ID Number: 617726Y6REV2, UDI (01)00840682118552(11)170900(21)REVVX1700090CN, Mfg. Lot/Serial Number REVVX1700090CN; System ID Number: 617726Y6REV1, UDI (01)00840682118552(11)160700(21)REVVX1600018CN, Mfg. Lot/Serial Number REVVX1600018CN; System ID Number: 336890APEX, UDI (01)00195278460271(11)240900(21)REV2X2400116CN, Mfg. Lot/Serial Number REV2X2400116CN; System ID Number: 704801APEX, UDI (01)00195278460271(11)241000(21)REV2X2400139CN, Mfg. Lot/Serial Number REV2X2400139CN; System ID Number: 336832APEX, UDI (01)001952
- Distribution
- Worldwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 19 Mar 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.