Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access…

Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Published
30 Apr 2025
Units affected
370 units
Agency
FDA
Country
United States

The hazard

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Brand / company
Beckman Coulter, Inc.
Codes / lots
UDI/DI 15099590732103, all units run with system software versions SW 1.20.0 and below
Distribution
Worldwide distribution.
Recall status
Ongoing