Fresenius Kabi USA, LLC recalls IVENIX INFUSION SYSTEM LVP Blood…

Fresenius Kabi USA, LLC recalls IVENIX INFUSION SYSTEM LVP Blood Products Administration Set

Published
18 Jun 2025
Units affected
14,280 units (724 cases)
Agency
FDA
Country
United States

The hazard

Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.

Brand / company
Fresenius Kabi USA, LLC
Codes / lots
Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015.
Distribution
US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.
Recall status
Ongoing