CheckMyRecalls

GE Medical Systems, LLC recalls SIGNA Architect and 3 more

Published
30 Jul 2025
Units affected
4 products
Agency
FDA
Country
United States

The hazard

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

SIGNA Architect

SIGNA Architect AIR

SIGNA PET/MR

Discovery MR750w 3.0T

Brand / company
GE Medical Systems, LLC
Codes / lots
GTIN: 00840682147095 00840682122702 00195278023643 00840682123440 | GTIN: 00195278283481 | GTIN: 00840682105378 | GTIN: 00840682103817 00195278229519
Distribution
Worldwide distribution - US Nationwide and the country of China.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 30 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.