Canon Medical System, USA, INC. recalls Alphenix INFX-8000F and 3 more
- Published
- 4 Feb 2026
- Units affected
- 4 products
- Agency
- FDA
- Country
- United States
The hazard
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Alphenix INFX-8000F, interventional fluoroscopic x-ray system
Alphenix INFX-8000H, interventional fluoroscopic x-ray system
Alphenix INFX-8000V, interventional fluoroscopic x-ray system
Alphenix INFX-8000C, interventional fluoroscopic x-ray system
- Brand / company
- Canon Medical System, USA, INC.
- Codes / lots
- serial numbers: W1C0582204, B2A1582208, W2A1052118, A3582035, A2A2362689, B2B18Y2813, W1D07Y2447, B4582113, W2A1042111, W1C0622270, A3592048, W4A11X2138, W2A09Y2060, B2A16Z2476, W2A09Y2066, W4A1122041, B2A1482032, B2A1482027, A2A2252435, W1D0722369, B2B15Y2247, W1B0542168, B2A1492049, W1D0722371, A2A2032031, A1C1972103, W4C1352375, B2B1862737, W1D0842499, W4A1212179, W4A11Z2171, W4A11Z2173, W1C0652298, B2A1692419, B2A1552170, A2A21Y2350, W4A1242226, B2B1952848, A1C19Z2180, B2A1662371, W1D0892555, W1D0822476, W1D06Z2356, B2A1682394, B2A17Y2619, B2A17Y2624, W1D0762406, A2A2172283, W4A11Y2155, A2A2272460, W1D0832487, B2B1822667, A1C1992146, W4A1112031, W4C1382409, A2A2292508, W4A13X2015, W4A1192130, B2A1712510, B2A1762562, W4A1142066, W1C0622274, A4552098, W4A12X2298, W1D0812460, A4542088, W1D0712363, A2A21X2348, W1C0552179, W4A1142065, B2A17Z2626, W4D13Y2444, W2A1042110, A2A20Z2159, W4A11Y2147, W1C0642289, B2B1852720, B2A1592219, W4B13Y2016, B2B1922843, B2B1822668, W4B12Y2312, A2A2052056, A2A2062062, A2A2072075, A2A2042041, A1C19Z2181, W2A1092163, A2A2012005, A2A20Y2147, A1C19Y2161, B2A1612289, W4A1112030, W2A0982042, B2B18X2802, B2A1712503, B2A1712499, B2A14Z2084. | serial numbers:
- Distribution
- Worldwide distribution - US Nationwide and the country of Dominican Republic.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 4 Feb 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.