CheckMyRecalls

Canon Medical System, USA, INC. recalls Alphenix Models INFX-8000C…

Canon Medical System, USA, INC. recalls Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a d…

Published
2 Sept 2026
Units affected
10 systems
Agency
FDA
Country
United States

The hazard

Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Brand / company
Canon Medical System, USA, INC.
Codes / lots
Model Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
Distribution
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 2 Sept 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.