Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System

Published
11 Feb 2026
Units affected
50
Agency
FDA
Country
United States

The hazard

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

CellMek SPS Sample Preparation System, REF: C44603, with software

Brand / company
Beckman Coulter, Inc.
Codes / lots
UDI-DI: 15099590750312. Software Version 2.3.90.0
Distribution
Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.
Recall status
Ongoing