Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System
- Published
- 11 Feb 2026
- Units affected
- 50
- Agency
- FDA
- Country
- United States
The hazard
Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
CellMek SPS Sample Preparation System, REF: C44603, with software
- Brand / company
- Beckman Coulter, Inc.
- Codes / lots
- UDI-DI: 15099590750312. Software Version 2.3.90.0
- Distribution
- Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.