Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A

Published
22 Apr 2026
Units affected
73
Agency
FDA
Country
United States

The hazard

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ARTIS Pheno VE30A and VE40A, Model 10849000

Brand / company
Siemens Medical Solutions USA, Inc
Codes / lots
UDI 04056869046877
Distribution
U.S. and O.U.S.
Recall status
Ongoing