Siemens Medical Solutions USA, Inc recalls ARTIS Pheno VE30A and VE40A
- Published
- 22 Apr 2026
- Units affected
- 73
- Agency
- FDA
- Country
- United States
The hazard
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
ARTIS Pheno VE30A and VE40A, Model 10849000
- Brand / company
- Siemens Medical Solutions USA, Inc
- Codes / lots
- UDI 04056869046877
- Distribution
- U.S. and O.U.S.
- Recall status
- Ongoing