CheckMyRecalls

Siemens Medical Solutions USA, Inc recalls ARTIS Icono biplane and 1…

Siemens Medical Solutions USA, Inc recalls ARTIS Icono biplane and 1 more

Published
10 Jun 2026
Units affected
2 products
Agency
FDA
Country
United States

The hazard

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system

ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system

Brand / company
Siemens Medical Solutions USA, Inc
Codes / lots
ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325 | UDI: 4056869046877
Distribution
US and Worldwide
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 10 Jun 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.