SIEMENS MEDICAL SOLUTIONS USA, INC recalls ARTIS One Angiographic…

SIEMENS MEDICAL SOLUTIONS USA, INC recalls ARTIS One Angiographic X-Ray System

Published
5 Feb 2025
Units affected
1140
Agency
FDA
Country
United States

The hazard

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

ARTIS One Angiographic X-Ray System

Brand / company
SIEMENS MEDICAL SOLUTIONS USA, INC
Codes / lots
Model Number 10848600
Distribution
23 Distributed in the US
Recall status
Ongoing