CheckMyRecalls

Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra (DE) and 14…

Siemens Medical Solutions USA, Inc recalls MAGNETOM Skyra (DE) and 14 more

Published
8 Oct 2025
Units affected
15 products
Agency
FDA
Country
United States

The hazard

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MAGNETOM Skyra (DE). Model Number: 10432915. 510(k) Numbers: K101347, K111242, K123510, K133435, K140253, K141977, K142515, K153343, K173592, K202014, K231560.

Biograph mMR. Model Number: 10433372.

MAGNETOM Verio. Model Number: 10276755.

MAGNETOM Lumina (DE). Model Number: 11344916.

MAGNETOM Spectra as the following: 1. MAGNETOM Spectra. Model Number: 10837643. 2. MAGNETOM Spectra (CN). Model Number: 10655588.

MAGNETOM Skyra Fit BioMatrix. Model Number: 11516217.

MAGNETOM Vida (DE). Model Number: 11060815. 510(k): K170396, K173617, K181433, K183254, K192924, K203443, K213693, K213805, K220425, K231560.

MAGNETOM Vida Fit. Model Number: 11410481.

MAGNETOM Skyra fit. Model Number: 10849580.

MAGNETOM Cima.X (DE). Model Number: 11647158.

MAGNETOM Verio Dot Upgrade. Model Number: 10684334.

MAGNETOM Connectom.X. Model Number: 11371480.

Brand / company
Siemens Medical Solutions USA, Inc
Codes / lots
Model Number: 10432915. UDI Numbers: (01)04056869006703(21)45046, (01)04056869006703(21)45115, (01)04056869006703(21)45398, (01)04056869006703(21)145760, (01)04056869006703(21)145811, (01)04056869006703(21)45549, (01)04056869006703(21)45834, (01)04056869006703(21)146099, (01)04056869006703(21)146110, (01)04056869006703(21)45047, (01)04056869006703(21)146093, (01)04056869006703(21)46066, (01)04056869006703(21)45788, (01)04056869006703(21)146082, (01)04056869006703(21)45124, (01)04056869006703(21)45185, (01)04056869006703(21)45332, (01)04056869006703(21)45466, (01)04056869006703(21)145390, (01)04056869006703(21)45732, (01)04056869006703(21)45918, (01)04056869006703(21)45380, (01)04056869006703(21)46182, (01)04056869006703(21)45735, (01)04056869006703(21)45233, (01)04056869006680(21)162026, (01)04056869006703(21)145008, (01)04056869006703(21)145473, (01)04056869006703(21)145095, (01)04056869006703(21)145664, (01)04056869006703(21)45200, (01)04056869006703(21)146098, (01)04056869006703(21)146086, (01)04056869006703(21)45805, (01)04056869006703(21)45812, (01)04056869006703(21)45397, (01)04056869006703(21)46264, (01)04056869006703(21)45499, (01)04056869006703(21)45749, (01)04056869006703
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 8 Oct 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.