Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade

Published
19 Nov 2025
Units affected
4 system (1 US; 3 OUS)
Agency
FDA
Country
United States

The hazard

There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

MAGNETOM Cima.X Upgrade. Model Number: 11689304.

Brand / company
Siemens Medical Solutions USA, Inc
Codes / lots
Model Number: 11689304. UDI: (01)04056869975245(21)237304. Serial Number: 237304.
Distribution
US distribution to California. International distribution to Australia, Germany, Great Britian.
Recall status
Ongoing