CheckMyRecalls

Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT…

Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit) and 3 more

Published
8 Jul 2026
Units affected
4 products
Agency
FDA
Country
United States

The hazard

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device

Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device

Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device

Brand / company
Baxter Healthcare Corporation
Codes / lots
UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 10887761981499; Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 00887761985018, Lot Numbers: All lot numbers distributed beginning on 4/28/2025 | UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 8 Jul 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.