Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and…

Baxter Healthcare Corporation recalls Baxter Duo-Vent Solution Set and Male Luer Lock Adapter

Published
19 Aug 2026
Units affected
24528 units
Agency
FDA
Country
United States

The hazard

Baxter received customer reports of air bubbles in the drip chamber and tubing when a pressure cuff is in use or when the tubing is flushed at the fully open position.

Classification: Class I — the most serious type: there is a reasonable chance this product could cause serious health problems or death.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product Code 1C8507; Solution Administration Sets

Brand / company
Baxter Healthcare Corporation
Codes / lots
UDI/DI 00085412014319, Lot Numbers: DR26C27062, exp. 30-MAR-2028; DR26D14031, exp. 15-Apr-2028; DR26E13055, exp. 14-May-2028
Distribution
US Nationwide. No OUS consignees.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 19 Aug 2026. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.