Ortho-Clinical Diagnostics, Inc. recalls Brand Name
- Published
- 9 Jul 2025
- Units affected
- 3 products
- Agency
- FDA
- Country
- United States
The hazard
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available software versions); Product Description: In vitro diagnostic chemistry analyzer.
Brand Name: VITROS Product Name: VITROS 4600 Chemistry System Model/Catalog Number: (1) 6802445, (2) 6900440 Software Version: version 3.2 to version 3.8.3 (all available software versions) Product Description: In vitro diagnostic chemistry analyzer.
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: (1) 6802413, (2) 6802915
- Brand / company
- Ortho-Clinical Diagnostics, Inc.
- Codes / lots
- VITROS XT 7600 Integrated System; Product Code: 6844461; UDI: 10758750012343; VITROS XT 7600 System - Certified; Product Code: 6272222; UDI: 10758750012343; Range of installed serial numbers:76000109-76002327; Total systems currently installed in the US: 1,019 Total systems currently installed outside the US: 939 Total systems installed worldwide (Product Quantity Distributed): 1,958 Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured, the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial num
- Distribution
- Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 9 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.