CheckMyRecalls

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Ca…

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products Ca Slides

Published
27 Aug 2025
Units affected
174,215 units
Agency
FDA
Country
United States

The hazard

Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;

Brand / company
Ortho-Clinical Diagnostics, Inc.
Codes / lots
Catalog Number: 145 0261; UDI-DI: 10758750009114; Affected GENs: 67 and above;
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
Recall status
Ongoing

Source: U.S. Food and Drug Administration, published 27 Aug 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.