Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products LAC…
Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products LAC Slides and 1 more
- Published
- 24 Sept 2025
- Units affected
- 2 products
- Agency
- FDA
- Country
- United States
The hazard
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
- Brand / company
- Ortho-Clinical Diagnostics, Inc.
- Codes / lots
- Catalog number 8150112; UDI# 10758750004454; Coating 0130 and above, this includes Lot numbers 3533-0130-2947, 3533-0130-2972, 3533-0130-3001, 3533-0130-3002, 3533-0130-3048, 3533-0130-3084, 3533-0131-5887, 3533-0131-5958, 3533-0131-5962, 3533-0131-6058 ; Expiration date range 01-Oct-2026 to 01-Jan-2027 | Catalog number 8433880; UDI# 110758750004911; Coating 0130 and above, this includes Lot numbers 3533-0130-2317, 3533-0130-2717, 3533-0130-2718, 3533-0130-2719, 3533-0130-4546, 3533-0130-4605, 3533-0131-4763, 3533-0131-6114, 3533-0131-6115, 3533-0131-6116, 3533-0131-6128; Expiration date range 01-Sep-2026 to 01-Jan-2027
- Distribution
- US Nationwide distribution.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 24 Sept 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.