Ortho-Clinical Diagnostics, Inc. recalls Brand Name and 2 more
- Published
- 23 Jul 2025
- Units affected
- 4 products
- Agency
- FDA
- Country
- United States
The hazard
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What to do
Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.
- Check the model and lot or serial number on the device label.
- Speak to your doctor or the clinic that supplied the device before stopping use.
Product details
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier II Model/Catalog Number: 8231474 Software Version: N/A Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component
VITROS Chemistry Products CHE Slides; Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides);
VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);
- Brand / company
- Ortho-Clinical Diagnostics, Inc.
- Codes / lots
- Lot Code: Product Code: 806 7324 (UDI: 10758750004317) Lots currently within expiry: V1665 - Exp.8/14/2025 B2245 - Exp 3/25/2026 Y2110 - Exp. 3/25/2026 E2828 - Exp. 2/3/2027 | Lot Code: Product Code: 823 1474 (UDI: 10758750004577) Lots currently within expiry: W1667- Exp. 8/14/2025 X1901 - Exp. 11/20/2025 A2114 - Exp. 3/25/2026 D2338 - Exp. 6/3/2026 F2830 - Exp. 2/3/2027 | Catalog Numbers: (1) 191 4605 (300 slides), (2) 800 4707 (90 slides); UDI-DI: (1) 10758750000241, (2) 10758750004225; Lot Numbers: All lots within expiry. | Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides); UDI-DI: (1) 10758750030293, (2) 10758750030309; Lot Numbers: All lots within expiry.
- Distribution
- Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
- Recall status
- Ongoing
Source: U.S. Food and Drug Administration, published 23 Jul 2025. CheckMyRecalls reproduces the official notice in summary; always confirm the details with the original notice.