CheckMyRecalls

Ortho-Clinical Diagnostics, INc. recalls Brand Name

Published
29 Jan 2025
Units affected
3 products
Agency
FDA
Country
United States

The hazard

The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.

Classification: Class II — this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What to do

Contact your doctor, clinic or the device maker to find out if your device is affected and what to do next. Do not stop using a life-supporting device without medical advice.

  • Check the model and lot or serial number on the device label.
  • Speak to your doctor or the clinic that supplied the device before stopping use.
Read the official FDA notice

Product details

Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component

Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: no, not a component

Brand / company
Ortho-Clinical Diagnostics, INc.
Codes / lots
Lot Code: VITROS XT 7600 Integrated System Product Code: 6844461 (New) UDI: 10758750012343 VITROS XT 7600 System (Certified) Product Code: 6272222 UDI: 10758750012343 Lot Code: VITROS XT 7600 Integrated System Product Code: 6904003 (China) UDI: 10758750035809 Range of installed serial numbers:76000001-76002199 Manufacture date range: July 2018 to present Distribution date range; July 2018 to present Total systems currently installed in the US: 994 Total systems currently installed outside the US: 899 Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
Recall status
Ongoing